| Error Code / Message | Meaning | Solution (per manual) | | :--- | :--- | :--- | | | Protein or fibrin buildup at the aperture. | Run "Clog Removal" cycle. If persists, perform "Aperture Soak" using Cleaner. | | "Fluid Leak Detected" | Internal sensor has triggered. | Check all tubing connections. Empty waste container. Call service if leak is visible. | | "WBC Interference" | High background noise. | Run two "Rinse" cycles. Check the lyse reagent for contamination. Run a background check with diluent. | | "RBC/PLT Noise" | Electrical interference or micro-clots. | Check grounding. Re-draw patient sample (possible clot in tube). | | "Reagent Empty" | One reagent is out or air is in the line. | Replace the reagent bottle, then run "Reagent Prime" three times. |
Proper installation is critical for ensuring the longevity and accuracy of the device. mindray bc 2800 vet user manual
Action Step: If the error persists after following the manual’s flowchart, contact Mindray technical support. Do not disassemble the fluidics system yourself. | Error Code / Message | Meaning |
Tell you how to (like HGB or clog errors) Explain the start-up or shut-down procedure Provide instructions for running a quality control sample Let me know which part of the manual you need. Manual BC 2800 Vet | PDF | Computer Keyboard - Scribd | | "Fluid Leak Detected" | Internal sensor has triggered